RecallRadar

FDA Device recall Z-1335-2021

BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring o…

On 2021-04-07 the FDA classified a Class II recall of BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring o… by Covidien, citing the capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to rem….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1335-2021
ClassificationClass II
Statusongoing
Initiated2021-02-26
Published2021-04-07
Last updated2026-09-13 11:56 UTC
CompanyCovidien
DistributionUS

Product

BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.

Reason

The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.

Identifiers

code_info
49349Q 49350Q 49351Q 49352Q 49799Q 49800Q 49801Q 49802Q 49809Q 49849Q 49850Q 49851Q 50392Q 50393Q 50492Q…49349Q 49350Q 49351Q 49352Q 49799Q 49800Q 49801Q 49802Q 49809Q 49849Q 49850Q 49851Q 50392Q 50393Q 50492Q 50493Q 50494Q 50495Q 50496Q 50497Q 50615Q 50616Q 50868Q 50869Q 50870Q 50871Q 50872Q 50873Q 50874Q 50875Q 50876Q 50877Q 50878Q 50879Q 50880Q 50881Q

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1335-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.