FDA Device recall Z-1329-2013
PROFILE-V MEDTOXScan Drugs of Abuse Test System
- FDA Device
- Class II
- terminated
- Published 2013-05-22
On 2013-05-22 the FDA classified a Class II recall of PROFILE-V MEDTOXScan Drugs of Abuse Test System by Medtox Diagnostics, citing review of validation information and customer complaints which indicated preliminary screen positive results for cocaine that were not positive on confirmatory testing.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1329-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-04-05 |
| Published | 2013-05-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Medtox Diagnostics |
| Distribution | US |
Product
PROFILE-V MEDTOXScan Drugs of Abuse Test System. For the rapid, qualitative detection of drugs of abuse in human urine,
Reason
Review of validation information and customer complaints which indicated preliminary screen positive results for cocaine that were not positive on confirmatory testing.
Identifiers
| code_info | Part No. 604018 - Lot No. 6821, 6822, 6823, 6824, 6825, 6831, 6832, 6841, 6851, 6852, 6853, 6861, 6862, 6863, 6864, 686…Part No. 604018 - Lot No. 6821, 6822, 6823, 6824, 6825, 6831, 6832, 6841, 6851, 6852, 6853, 6861, 6862, 6863, 6864, 6865, 6866, 6867, 6868, 6871, 6881, 6891, 6892, 6901, 6902, 6903, 6904, 6911, 6912, 6921, 6922, 6931, 6942; Part No. 604019 - Lot No. 6321, 6331, 6341, 6351; Part No. 604020 - Lot No. 6361, 6371; Part No. 604022 - Lot No. 6551, 6552, 6561, 6562, 6563, 6564, 6571, 6581, 6582, 6583, 6584, 6591, 6592, 6593, 6601, 6602, 6603, 6611, 6612, 6621, 6622, 6623, 6631, 6632, 6641, 6651, 6652, 6661, 6671, 6672, 6681; Part No. 604026 - Lot No. 6041, 6051; Part No. 604028 - Lot No. 6051, 6061, 6071, 6081, 6091, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1329-2013%22