Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG 1024, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intende…

On 2019-05-22 the FDA classified a Class II recall of Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG 1024, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intende… by Pentax, citing incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the con….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1326-2019
ClassificationClass II
Statusterminated
Initiated2018-12-03
Published2019-05-22
Last updated2026-09-13 11:55 UTC
CompanyPentax
DistributionCA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA

Product

Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG 1024, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.

Reason

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Identifiers

code_infoLot/Serial numbers: 06152018-01, 07172018-02, 07022018-04, 08212018-01, and 08102018-03.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1326-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.