FDA Device recall Z-1325-2020
Abutment Screw Retrieval Instrument RP/WP/6.0 - Product Usage: The instruments for abutment screw retrieval can be used if an abutment scre…
- FDA Device
- Class II
- terminated
- Published 2020-02-26
On 2020-02-26 the FDA classified a Class II recall of Abutment Screw Retrieval Instrument RP/WP/6.0 - Product Usage: The instruments for abutment screw retrieval can be used if an abutment scre… by Nobel Biocare Usa, citing due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1325-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-03-14 |
| Published | 2020-02-26 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Nobel Biocare Usa |
| Distribution | US |
Product
Abutment Screw Retrieval Instrument RP/WP/6.0 - Product Usage: The instruments for abutment screw retrieval can be used if an abutment screw is broken and a part of the screw is left inside the implant screw channel. If the implant thread needs to be cleaned before placing a new screw, a screw tap repair can be used. The instruments are intended for both simple and advanced treatments.
Reason
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
Identifiers
| code_info | (01)07332747084092(10)73562 Catalog# 37504 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1325-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.