RecallRadar

FDA Device recall Z-1319-2026

Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH

On 2026-02-18 the FDA classified a Class I recall of Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH by Medline Industries, citing these lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, o….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1319-2026
ClassificationClass I
Statusongoing
Initiated2025-12-22
Published2026-02-18
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;

Reason

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Identifiers

code_info
BD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP25080…BD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP250808; BD710FJ282RRH UDI-DI 10197344044054 Lots EP250429 EP250808; BD710FJ282CRH UDI-DI (case) 40197344044048 (ea) 10197344044047 Lots EP250429 EP250808

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1319-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.