FDA Device recall Z-1319-2026
Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH
- FDA Device
- Class I
- ongoing
- Published 2026-02-18
On 2026-02-18 the FDA classified a Class I recall of Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH by Medline Industries, citing these lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, o….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1319-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-12-22 |
| Published | 2026-02-18 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medline Industries |
| Distribution | US |
Product
Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;
Reason
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Identifiers
| code_info | BD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP25080…BD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP250808; BD710FJ282RRH UDI-DI 10197344044054 Lots EP250429 EP250808; BD710FJ282CRH UDI-DI (case) 40197344044048 (ea) 10197344044047 Lots EP250429 EP250808 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1319-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.