RecallRadar

FDA Device recall Z-1318-2018

Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position…

On 2018-04-18 the FDA classified a Class II recall of Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position… by Reliance Medical Products, citing product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1318-2018
ClassificationClass II
Statusterminated
Initiated2017-12-06
Published2018-04-18
Last updated2026-09-13 11:54 UTC
CompanyReliance Medical Products
DistributionUS

Product

Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position ophthalmological examination and diagnostic equipment, is to store and position for examination a device commonly known as a refractor.

Reason

Product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing

Identifiers

code_infoModel Number: 525004, 05 and 525006, 07

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1318-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.