RecallRadar

FDA Device recall Z-1317-2018

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LO…

On 2018-04-18 the FDA classified a Class II recall of D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LO… by Smith And Nephew, citing one lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1317-2018
ClassificationClass II
Statusterminated
Initiated2018-03-07
Published2018-04-18
Last updated2026-09-13 11:54 UTC
CompanySmith And Nephew
DistributionPR

Product

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Reason

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

Identifiers

code_infoLot Number 17GM04522

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.