FDA Device recall Z-1317-2013
PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Elec…
- FDA Device
- Class II
- terminated
- Published 2013-05-22
On 2013-05-22 the FDA classified a Class II recall of PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Elec… by Conmed, citing conMed was notified by the contract manufacturer that their wire/connector supplier reported damage to the connector tooling which could pinch or compress the lead wire filaments….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1317-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-05-03 |
| Published | 2013-05-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Conmed |
| Distribution | US |
Product
PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Electrodes with Carbon Leadwires and Medtronic(R) Connector are intended to be used in external pacing, defibrillation and monitoring applications
Reason
ConMed was notified by the contract manufacturer that their wire/connector supplier reported damage to the connector tooling which could pinch or compress the lead wire filaments and potentially render the device incapable of delivering the appropriate energy or shock to the patient and/or could cause potential injury to user or patient.
Identifiers
| code_info | Lot Code Y092012-21, Manufacture Date 10/10/2012; Lot Code Y102612-12, Manufacture Date 11/17/2012; Lot Code Y110912…Lot Code Y092012-21, Manufacture Date 10/10/2012; Lot Code Y102612-12, Manufacture Date 11/17/2012; Lot Code Y110912-20, Manufacture Date 11/30/2012; Lot Code Y112912-10, Manufacture Date 12/22/2012; Lot Code Y121312-19, Manufacture Date 01/03/2013; Lot Code Y011013-13, Manufacture Date 01/29/2013; Lot Code Y012513-02, Manufacture Date 02/12/2013; Lot Code Y020713-05, Manufacture Date 02/26/2013; Lot Code Y021413-03, Manufacture Date 03/07/2013; Lot Code Y030113-04, Manufacture Date 03/23/2013 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1317-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.