FDA Device recall Z-1315-2015
Liquid Mouse Monoclonal Antibody CD15 used is laboratories to perform immunohistochemical (IHC) staining
- FDA Device
- Class II
- terminated
- Published 2015-04-01
On 2015-04-01 the FDA classified a Class II recall of Liquid Mouse Monoclonal Antibody CD15 used is laboratories to perform immunohistochemical (IHC) staining by Leica Microsystems, citing the staining intensity decreases over the shelf life.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1315-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-12-16 |
| Published | 2015-04-01 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Leica Microsystems |
| Distribution | US |
Product
Liquid Mouse Monoclonal Antibody CD15 used is laboratories to perform immunohistochemical (IHC) staining. Product Usage: NCL-L-CD15 is intended for the qualitative identification by light microscopy of CD15 molecules in paraffin sections. The clinical interpretation of any staining or its absence should be complemented by morphological studies using proper controls and should be evaluated within the context of the patients clinical history and other diagnostic tests by a qualified pathologist.
Reason
The staining intensity decreases over the shelf life.
Identifiers
| code_info | Product Code: NCL-L-CD15; Lot Numbers and Expiration Dates: Lot 6022274, Expiry 2015-04; Lot 6024628, Expiry 2015-08; Lot 6030581, Expiry 2016-07 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1315-2015%22