FDA Device recall Z-1314-2014

AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet

On 2014-04-09 the FDA classified a Class II recall of AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet by Allesee Orthodontic Appliances, citing the Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this colla….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1314-2014
ClassificationClass II
Statusterminated
Initiated2014-02-24
Published2014-04-09
Last updated2026-09-13 11:53 UTC
CompanyAllesee Orthodontic Appliances
DistributionUS

Product

AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet. Removable functional orthodontic appliances that are to modify the growth of the jaws in a prescribed growth pattern.

Reason

The Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this collar, an over-tightened lock nut may stop the piston assembly from moving freely, subsequently preventing intended movement of the teeth.

Identifiers

code_infoAll Devices/Codes Manufactured from August 28, 2013 to November 6, 2013