RecallRadar

FDA Device recall Z-1310-2019

MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Val…

On 2019-05-29 the FDA classified a Class I recall of MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Val… by Integra Lifesciences, citing the firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1310-2019
ClassificationClass I
Statusterminated
Initiated2019-04-08
Published2019-05-29
Last updated2026-09-13 11:55 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Valve w MoniTorr 10110 10-110 CSF Drainage System used with Pole Mount System MoniTorr 10140 10-140 CSF Drainage System w/Vinyl Measuring Strip MoniTorr 10150 10-150 CSF Drainage System Simple Bag and Line MoniTorr 10110A 10-110A CSF Drainage System w/o Manifold Stopcock MoniTorr SP0017 SP0017 Special EVD 10-110 w/o Y Site Latex Free Sites MoniTorr SP0042 SP0042 Special EVD 10-100 w/o Y Site MoniTorr SP0090 SP0090 Special EVD 10-110 w/o Y Site MoniTorr SP0164 SP0164 Special EVD 10-140 w/Y Site & Stopcock Reverse MoniTorr SP0236 SP0236 INS 1100 WITH NEEDLE AT Y-ACCESS AND NEEDLELESS SI Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, closed system for the drainage of cerebrospinal fluid (CSF) from ventricles of the brain or the lumbar subarachnoid space to a drainage bag. The system may be used with a pole mounted assembly that allows for simple, quick and accurate alignment with the patient and secure positive or negative pressure level setting. The system also has been designed to provide for ease of patient transport through a compact design and antimicrobial hydrophobic vent feature that resists occlusion. Indications: The MoniTorr ICP system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF inpatients with infected CSF shunts, and to monitor ICP.

Reason

The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.

Identifiers

code_infoAll lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1310-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.