FDA Device recall Z-1309-2026
Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M1500…
- FDA Device
- Class II
- ongoing
- Published 2026-02-11
On 2026-02-11 the FDA classified a Class II recall of Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M1500… by Etac A S, citing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1309-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-17 |
| Published | 2026-02-11 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Etac A S |
| Distribution | US |
Product
Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. Component: The deformation occurs between the bolt and the fork in the joint connecting the lifting bar and the arm on the Molift Mover.
Reason
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Identifiers
| code_info | (1) Model/Catalog Number: M1501; UDI-DI: TBD; Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 6…(1) Model/Catalog Number: M1501; UDI-DI: TBD; Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 621102, 611505, 611504, 611503, 611502, 611501, 601008, 611507, 601007, 601006, 601001, 621103, 601003, 601004; (2) Model/Catalog Number: M1502; UDI-DI: TBD; Serial Numbers: 611409, 611408, 611407, 611406, 611405, 611404, 611403, 611402, 611401, 611412, 611411, 611410; (3) Model/Catalog Number: M1504; UDI-DI: TBD; Serial Numbers: 600902, 600901, 611303, 611302, 611301; (4) Model/Catalog Number: M15050; UDI-DI: TBD; Serial Numbers: 586003, 586001, 585905, 585903, 585902, 585901, 598605, 586101, 585904, 603602, 591401, 586002, 586006, 586005, 586004, 598608, 598607, 585906, 591601, 598606, 598604, 598602, 598601, 596501, 603601, 598609, 598603; (5) Model/Catalog Number: M15151; UDI-DI: TBD; Serial Numbers: 599703, 599702, 599701; (6) Model/Catalog Number: M15001; UDI-DI: TBD; Serial Numbers: 584401; (7) Model/Catalog Number: M15002; UDI-DI: TBD; Serial Numbers: 5915 |
|---|