RecallRadar

FDA Device recall Z-1309-2026

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M1500…

On 2026-02-11 the FDA classified a Class II recall of Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M1500… by Etac A S, citing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1309-2026
ClassificationClass II
Statusongoing
Initiated2025-11-17
Published2026-02-11
Last updated2026-09-13 11:58 UTC
CompanyEtac A S
DistributionUS

Product

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. Component: The deformation occurs between the bolt and the fork in the joint connecting the lifting bar and the arm on the Molift Mover.

Reason

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Identifiers

code_info
(1) Model/Catalog Number: M1501; UDI-DI: TBD; Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 6…(1) Model/Catalog Number: M1501; UDI-DI: TBD; Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 621102, 611505, 611504, 611503, 611502, 611501, 601008, 611507, 601007, 601006, 601001, 621103, 601003, 601004; (2) Model/Catalog Number: M1502; UDI-DI: TBD; Serial Numbers: 611409, 611408, 611407, 611406, 611405, 611404, 611403, 611402, 611401, 611412, 611411, 611410; (3) Model/Catalog Number: M1504; UDI-DI: TBD; Serial Numbers: 600902, 600901, 611303, 611302, 611301; (4) Model/Catalog Number: M15050; UDI-DI: TBD; Serial Numbers: 586003, 586001, 585905, 585903, 585902, 585901, 598605, 586101, 585904, 603602, 591401, 586002, 586006, 586005, 586004, 598608, 598607, 585906, 591601, 598606, 598604, 598602, 598601, 596501, 603601, 598609, 598603; (5) Model/Catalog Number: M15151; UDI-DI: TBD; Serial Numbers: 599703, 599702, 599701; (6) Model/Catalog Number: M15001; UDI-DI: TBD; Serial Numbers: 584401; (7) Model/Catalog Number: M15002; UDI-DI: TBD; Serial Numbers: 5915

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1309-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.