FDA Device recall Z-1308-2026
Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402…
- FDA Device
- Class II
- ongoing
- Published 2026-02-11
On 2026-02-11 the FDA classified a Class II recall of Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402… by Etac A S, citing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1308-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-17 |
| Published | 2026-02-11 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Etac A S |
| Distribution | US |
Product
Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;
Reason
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Identifiers
| code_info | (1) Model/Catalog Number: M14000; UDI-DI: 05707995032270; Serial Numbers: 616606, 616604, 591003, 591002, 590906, 59090…(1) Model/Catalog Number: M14000; UDI-DI: 05707995032270; Serial Numbers: 616606, 616604, 591003, 591002, 590906, 590901, 590904, 590903, 590902, 591001, 604101, 604005, 591703, 591007, 591006, 591005, 591004, 615105, 615104, 615103, 615102, 615101, 604004, 604003, 604002, 604001, 614202, 614201, 591702, 591701, 616603, 615106, 590905; (2) Model/Catalog Number: M1401; UDI-DI: 05707995032270; Serial Numbers: 593302, 593301, 593303, 612507, 612506, 612508, 612505, 612503, 620702, 612504, 612502, 612501, 593304, 593305, 620703, 620701; (3) Model/Catalog Number: M1402; UDI-DI: 05707995032270; Serial Numbers: 612601; (4) Model/Catalog Number: M14050; UDI-DI: 05707995032270; Serial Numbers: 590801, 596701, 585701, 584501, 591801, 604405, 604406, 604402, 604404, 604403, 604401, 604303, 604302, 604301; (5) Model/Catalog Number: M14150; UDI-DI: 05707995032270; Serial Numbers: 585811, 585810, 585809, 585808, 585806, 585805, 585804, 585803, 585801, 596604, 596603, 596602, 596601, 596611, 5966 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1308-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.