RecallRadar

FDA Device recall Z-1304-2018

Dynavox T10 Speech Generating Device ( including Batteries)

On 2018-04-11 the FDA classified a Class II recall of Dynavox T10 Speech Generating Device ( including Batteries) by Dynavox Systems, citing the battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1304-2018
ClassificationClass II
Statusterminated
Initiated2017-12-14
Published2018-04-11
Last updated2026-09-13 11:54 UTC
CompanyDynavox Systems
Distributionn/a

Product

Dynavox T10 Speech Generating Device ( including Batteries)

Reason

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Identifiers

code_info
TS1308 TS1411 Part Numbers: T10: 750394 Date Codes are part of the device s unique Serial Number and represent…TS1308 TS1411 Part Numbers: T10: 750394 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1304-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.