FDA Device recall Z-1302-2025
GE Healthcare Revolution CT, System, X-ray, Tomography, Computed
- FDA Device
- Class II
- ongoing
- Published 2025-03-19
On 2025-03-19 the FDA classified a Class II recall of GE Healthcare Revolution CT, System, X-ray, Tomography, Computed by Ge Medical Systems, citing GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1302-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-02-18 |
| Published | 2025-03-19 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Ge Medical Systems |
| Distribution | n/a |
Product
GE Healthcare Revolution CT, System, X-ray, Tomography, Computed
Reason
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Identifiers
| code_info | System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN…System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 480342APEX, UDI (01)00840682146616(11)220900(21)REV2A2200072CN, Mfg. Lot/Serial Number REV2A2200072CN; System ID Number 480342APEX3, UDI (01)00195278460271(11)240600(21)REV2X2400067CN, Mfg. Lot/Serial Number REV2X2400067CN; System ID Number 949250APEX, UDI (01)00195278460271(11)230500(21)REV2X2300008CN, Mfg. Lot/Serial Number REV2X2300008CN; System ID Number 858657APEX, UDI (01)00840682118552(11)160900(21)REVVX1600034CN, Mfg. Lot/Serial Number REVVX1600034CN; System ID Number 310423CT, UDI (01)00840682118552(11)211200(21)REVVX2100047CN, Mfg. Lot/Serial Number REVVX2100047CN; System ID Number 707257APEX1, UDI (01)00195278460271(11)231000(21)REV2X2300077CN, Mfg. Lot/Serial Number REV2X2300077CN; System ID Number 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number 408559CTREVO, UDI To be provided, Mfg. Lot/Ser |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1302-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.