FDA Device recall Z-1301-2018

Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Bi…

On 2018-04-11 the FDA classified a Class II recall of Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Bi… by The Binding Site Group, citing the product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1301-2018
ClassificationClass II
Statusterminated
Initiated2018-03-09
Published2018-04-11
Last updated2026-09-13 11:54 UTC
CompanyThe Binding Site Group
DistributionCA

Product

Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Birmingham, U.K. The Optilite Beta-2-Microglobulin (¿2M) Urine kit is intended for the quantitative in-vitro measurement of beta-2-microglobulin in urine using the Binding Site Optilite analyser to aid in the diagnosis of active rheumatoid arthritis and kidney disease. This test should be used in conjunction with other laboratory and clinical findings.

Reason

The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.

Identifiers

code_infoLot 407460, UDI 05051700017688

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.