FDA Device recall Z-1294-2019

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washi…

On 2019-05-15 the FDA classified a Class II recall of Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washi… by Medica, citing wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1294-2019
ClassificationClass II
Statusterminated
Initiated2019-02-06
Published2019-05-15
Last updated2026-09-13 11:55 UTC
CompanyMedica
DistributionUS

Product

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.

Reason

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

Identifiers

code_infoLot #s 18066, 18133, 18235, 18317