RecallRadar

FDA Device recall Z-1286-2021

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)108468…

On 2021-03-31 the FDA classified a Class II recall of Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)108468… by Biosense Webster, citing there is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1286-2021
ClassificationClass II
Statuscompleted
Initiated2021-02-26
Published2021-03-31
Last updated2026-09-13 11:56 UTC
CompanyBiosense Webster
Distributionn/a

Product

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016260 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Reason

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Identifiers

code_infoAll Lots