FDA Device recall Z-1286-2015
CADD" Medication Cassette Reservoir, 50mL, REF 21-7001-24 & 21-7301-24, Sterile EO, Rx only
- FDA Device
- Class II
- terminated
- Published 2015-03-25
On 2015-03-25 the FDA classified a Class II recall of CADD" Medication Cassette Reservoir, 50mL, REF 21-7001-24 & 21-7301-24, Sterile EO, Rx only by Smiths Medical Asd, citing smiths Medical has become aware of an issue with specific lots of 50mL CADD" Medication Cassette Reservoirs (Cassette). Some Cassettes may leak at the sealing area of the pump t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1286-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-02-02 |
| Published | 2015-03-25 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Smiths Medical Asd |
| Distribution | US |
Product
CADD" Medication Cassette Reservoir, 50mL, REF 21-7001-24 & 21-7301-24, Sterile EO, Rx only.
Reason
Smiths Medical has become aware of an issue with specific lots of 50mL CADD" Medication Cassette Reservoirs (Cassette). Some Cassettes may leak at the sealing area of the pump tube and medication bag. Smiths Medical has received no reports of serious injury or death related to this issue.
Identifiers
| code_info | 21-7001-24: Lots 14X297 and 14X323 21-7301-24 (Flow Stop): Lot 14X324 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1286-2015%22