FDA Device recall Z-1282-2019

Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices

On 2019-05-15 the FDA classified a Class II recall of Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices by Medtronic, citing the application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiv….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1282-2019
ClassificationClass II
Statusterminated
Initiated2018-07-04
Published2019-05-15
Last updated2026-09-13 11:55 UTC
CompanyMedtronic
Distributionn/a

Product

Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

Reason

The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.

Identifiers

code_infoUDI - (01)00643169693142(10)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.