FDA Device recall Z-1281-2019

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomog…

On 2019-05-29 the FDA classified a Class II recall of NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomog… by Neusoft Medical Systems, citing software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system wh….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1281-2019
ClassificationClass II
Statusterminated
Initiated2018-10-22
Published2019-05-29
Last updated2026-09-13 11:55 UTC
CompanyNeusoft Medical Systems
DistributionUS

Product

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Identifiers

code_infoSeries numbers N16E180037E; N16E180047E

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1281-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.