RecallRadar

FDA Device recall Z-1281-2016

CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray im…

On 2016-04-06 the FDA classified a Class II recall of CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray im… by Siemens Medical Solutions Usa, citing siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system mainten….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1281-2016
ClassificationClass II
Statusterminated
Initiated2016-02-16
Published2016-04-06
Last updated2026-09-13 11:54 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

Reason

Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.

Identifiers

code_infoModel Number: 10308191

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1281-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.