FDA Device recall Z-1281-2016
CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray im…
- FDA Device
- Class II
- terminated
- Published 2016-04-06
On 2016-04-06 the FDA classified a Class II recall of CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray im… by Siemens Medical Solutions Usa, citing siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system mainten….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1281-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-02-16 |
| Published | 2016-04-06 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Reason
Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.
Identifiers
| code_info | Model Number: 10308191 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1281-2016%22