Prismaflex Control Unit

On 2018-04-11 the FDA classified a Class II recall of Prismaflex Control Unit by Baxter Healthcare, citing firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Addit….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1280-2018
ClassificationClass II
Statusterminated
Initiated2017-02-15
Published2018-04-11
Last updated2026-09-13 11:54 UTC
CompanyBaxter Healthcare
DistributionUS

Product

Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. *Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated.

Reason

Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.

Identifiers

code_infoProduct code: 6023014700, 107493, 113081, 115269; Serial Number: All; Affected Software Versions: SW 4.1x, 5.10, 6.10, 7.xx, 8.00