RecallRadar

FDA Device recall Z-1276-2018

Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA…

On 2018-04-11 the FDA classified a Class II recall of Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA… by Phadia Ab, citing we want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejectin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1276-2018
ClassificationClass II
Statusterminated
Initiated2017-11-20
Published2018-04-11
Last updated2026-09-13 11:54 UTC
CompanyPhadia Ab
DistributionCA, GA, IN, MA, ME, MS, NC, NJ, OR, TN, TX, UT, VA

Product

Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall is for any Phadia Prime software version up to and including the current version, 2.1.4.

Reason

We want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA assay (the OK function may be used for rejecting single tests and dilution of samples in accordance with product DfU).

Identifiers

code_info
Phadia Prime software 12-4101-00 Phadia 250 12-3900-01 Listed software: EliA dsDNA Well 14-5500-01 EliA U1RNP Wel…Phadia Prime software 12-4101-00 Phadia 250 12-3900-01 Listed software: EliA dsDNA Well 14-5500-01 EliA U1RNP Well 14-5501-01 EliA Sm Well 14-5502-01 EliA Ro Well 14-5503-01 EliA La Well 14-5504-01 EliA CENP Well 14-5505-01 EliA Scl-70 Well 14-5506-01 EliA Jo-1 Well 14-5507-01 EliA Symphony Well 14-5508-01 EliA RNP70 Well 14-5511-01 EliA GBM Well 14-5514-01 EliA CCP Well 14-5515-01 EliA Celikey IgA Well 14-5517-01 EliA Celikey IgG Well 14-5518-01 EliA Gliadin IgA Well 14-5519-01 EliA Gliadin IgG Well 14-5520-01 EliA Cardiolipin IgA Well 14-5528-01 EliA Cardiolipin IgG Well 14-5529-01 EliA Cardiolipin IgM Well 14-5530-01 EliA B2-Glycoprotein I IgA Well 14-5531-01 EliA B2-Glycoprotein I IgG Well 14-5532-01 EliA B2-Glycoprotein I IgM Well 14-5533-01 EliA PR3S Well 14-5536-01 EliA MPOS Well 14-5537-01 EliA GliadinDP IgA Well 14-5538-01 EliA GliadinDP IgG Well 14-5539-01 EliA RF IgM Well 14-5600-01 EliA RF IgA Well 14-5601-01 EliA SmDP Well 14-5624-01 EliA