FDA Device recall Z-1274-2025
Osstem Implant System - Abutment ET Multi Abutment Size: 4.8D 3.0G/H- Abutment is intended for use with a dental implant to provide suport…
- FDA Device
- Class II
- ongoing
- Published 2025-03-12
On 2025-03-12 the FDA classified a Class II recall of Osstem Implant System - Abutment ET Multi Abutment Size: 4.8D 3.0G/H- Abutment is intended for use with a dental implant to provide suport… by Hiossen, citing patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1274-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-01-27 |
| Published | 2025-03-12 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Hiossen |
| Distribution | US |
Product
Osstem Implant System - Abutment ET Multi Abutment Size: 4.8D 3.0G/H- Abutment is intended for use with a dental implant to provide suport for prosthetic restorations such as crowns, bridges, or overdentures. Model/Catalog Number: ETMTA503RV1
Reason
Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
Identifiers
| code_info | Lot Code: UDI-DI: 00810019241529 Lot# H4E21D092 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1274-2025%22