RecallRadar

FDA Device recall Z-1273-2020

TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone

On 2020-02-26 the FDA classified a Class II recall of TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone by Smith And Nephew, citing due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1273-2020
ClassificationClass II
Statusterminated
Initiated2019-12-30
Published2020-02-26
Last updated2026-09-13 11:55 UTC
CompanySmith And Nephew
DistributionUS

Product

TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.

Reason

Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.

Identifiers

code_infoBatch numbers: 2010547, 2025137, 2036695, 50763859, & 50773071. Part Number 72200796. UDI 10600147.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1273-2020%22