FDA Device recall Z-1273-2020
TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone
- FDA Device
- Class II
- terminated
- Published 2020-02-26
On 2020-02-26 the FDA classified a Class II recall of TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone by Smith And Nephew, citing due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1273-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-12-30 |
| Published | 2020-02-26 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Smith And Nephew |
| Distribution | US |
Product
TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.
Reason
Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.
Identifiers
| code_info | Batch numbers: 2010547, 2025137, 2036695, 50763859, & 50773071. Part Number 72200796. UDI 10600147. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1273-2020%22