AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imagi…

On 2014-04-02 the FDA classified a Class II recall of AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imagi… by Siemens Medical Solutions Usa, citing certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop ou….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1272-2014
ClassificationClass II
Statusterminated
Initiated2014-02-07
Published2014-04-02
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason

Certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system.

Identifiers

code_info
Model nos. 7412807, 7727717, 7728392 with serial numbers: 35020, 34999,35098 35113 35042 35061 35064 35073 35014…Model nos. 7412807, 7727717, 7728392 with serial numbers: 35020, 34999,35098 35113 35042 35061 35064 35073 35014 35069 35053 35055 35056 35054 35008 35057 35044 35018 35097 35108 35046 35048 35050 35045 35004 35019 46004 46005 40000 40009 40011 40006 40007 40010 44016 44005 44011