RecallRadar

FDA Device recall Z-1270-2015

GE Healthcare Engstrom Carestation, Engstrom Pro, Cart Only S/5 EC. GE model numbers 1505-8538-000, 1505-9000-000, and 1505-9003-000

On 2015-03-25 the FDA classified a Class II recall of GE Healthcare Engstrom Carestation, Engstrom Pro, Cart Only S/5 EC. GE model numbers 1505-8538-000, 1505-9000-000, and 1505-9003-000 by Ge Healthcare, citing potential failure of the caster mounting hardware, which may result in the ventilator falling/tipping (overbalancing).

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1270-2015
ClassificationClass II
Statusterminated
Initiated2015-01-27
Published2015-03-25
Last updated2026-09-13 11:53 UTC
CompanyGe Healthcare
DistributionUS

Product

GE Healthcare Engstrom Carestation, Engstrom Pro, Cart Only S/5 EC. GE model numbers 1505-8538-000, 1505-9000-000, and 1505-9003-000. Ventilators.

Reason

Potential failure of the caster mounting hardware, which may result in the ventilator falling/tipping (overbalancing).

Identifiers

code_info
Mfg Lot or Serial # CBCP02195 CBCP02196 CBCP02201 CBCP01764 CBCP01766 CBCP02109 CBCP01273 CBCP01409 CBCP01506…Mfg Lot or Serial # CBCP02195 CBCP02196 CBCP02201 CBCP01764 CBCP01766 CBCP02109 CBCP01273 CBCP01409 CBCP01506 CBCP01528 CBCP01533 CBCP01538 CBCP01554 CBCP01629 CBCP01630 CBCP01631 CBCP01632 CBCP01635 CBCP01636 CBCP01638 CBCP01639 CBCP01641 CBCP01642 CBCP01379 CBCP01380 CBCP01400 CBCP01408 CBCP01411 CBCP01417 CBCP01419 CBCP01420 CBCP01423 CBCP01425 CBCP01489 CBCP01490 CBCP01491 CBCP01494 CBCP01496 CBCP01501 CBCP01530 CBCP01540 CBCP01541 CBCP01562 CBCP01615 CBCP01619 CBCP01721 CBCP01724 CBCP01725 CBCP01726 CBCP01727 CBCP01733 CBCP01748 CBCP01116 CBCP01383 CBCP01402 CBCP01403 CBCP01429 CBCP01497 CBCP01560 CBCP01579 CBCP01580 CBCP01623 CBCP01660 CBCP01661 CBCP01704 CBCP01708 CBCP01723 CBCP01728 CBCP01730 CBCP01731 CBCP01734 CBCP01736 CBCP01737 CBCP01740 CBCP01744 CBCP01745 CBCP01746 CBCP01749 CBCP01750 CBCP01751 CBCP01755 CBCP01084 CBCP01092 CBCP01122 CBCP01248 CBCP01250 CBCP01254 CBCP01255 CBCP01258

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1270-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.