RecallRadar

FDA Device recall Z-1267-2025

MIM software

On 2025-03-12 the FDA classified a Class II recall of MIM software by Mim Software, citing in situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software ver….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1267-2025
ClassificationClass II
Statusongoing
Initiated2025-02-19
Published2025-03-12
Last updated2026-09-13 11:57 UTC
CompanyMim Software
DistributionUS

Product

MIM software; System, Image Processing, Radiological

Reason

In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.

Identifiers

code_info
Software versions: 1. MIM Anyware 1.1, UDI/DI Not Applicable 2. MIM 7.2.0, UDI/DI 00850009343192; 3. MIM 7.2.1, UDI/…Software versions: 1. MIM Anyware 1.1, UDI/DI Not Applicable 2. MIM 7.2.0, UDI/DI 00850009343192; 3. MIM 7.2.1, UDI/DI 00850009343192; 4. MIM 7.2.2, UDI/DI 00850009343192; 5. MIM 7.2.3, UDI/DI 00850009343192; 6. MIM 7.2.4, UDI/DI 00850009343192; 7. MIM 7.2.5, UDI/DI 00850009343192; 8. MIM 7.2.6, UDI/DI 00850009343192

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1267-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.