On 2015-03-18 the FDA classified a Class III recall of Carton: Richard-Allan Scientific CYTO-STAIN by Richard Allan Scientific, citing the case was correctly labeled as REF 7501R, however the gallon bottle was incorrectly labeled as REF 7501G Cyto-Stain G. The bottle label has the correct lot number of 313833 and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1264-2015 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2015-02-19 |
| Published | 2015-03-18 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Richard Allan Scientific |
| Distribution | n/a |
Product
Carton: Richard-Allan Scientific CYTO-STAIN; REF 7501R; Quantity: 1x1 Gallon; Size: 1 Gallon/3.8L; LOT 313833; 11/2016; Bottle: Richard-Allan Scientific CYTO-STAIN G; REF 7501G; 1 Gallon/3.8L; Thermo Scientific.
Reason
The case was correctly labeled as REF 7501R, however the gallon bottle was incorrectly labeled as REF 7501G Cyto-Stain G. The bottle label has the correct lot number of 313833 and EXP date 11/2016. The gallon bottle contains the correct intended contents of 7501R: Cyto-Stain R Stain.
Identifiers
| code_info | Lot: 313833 Expiration date: 11/2016 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1264-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.