RecallRadar

FDA Device recall Z-1262-2016

The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmace…

On 2016-04-06 the FDA classified a Class II recall of The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmace… by Siemens Medical Solutions Usa, citing we received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1262-2016
ClassificationClass II
Statusterminated
Initiated2016-02-26
Published2016-04-06
Last updated2026-09-13 11:54 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body

Reason

We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change

Identifiers

code_info
10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109,…10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109, 05991117, 05992099, 07324119, 07324135, 07324143, 07324150, 07760809, 07760932, 07761161, 07823920, 07823938, 07823946, 07823953, 07823961, 07823979, 10151531, 10151532, 10520745

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1262-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.