RecallRadar

FDA Device recall Z-1262-2014

Natural-Knee¿ II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA…

On 2014-03-26 the FDA classified a Class II recall of Natural-Knee¿ II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA… by Zimmer, citing A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was adva….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1262-2014
ClassificationClass II
Statusterminated
Initiated2014-02-25
Published2014-03-26
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

Natural-Knee¿ II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA TIB RT SZ0/16M ALL POLY ULTRA TIB LT SZ1 ALL POLY ULTRA TIB RT SZ1 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB RT SZ2 ALL POLY ULTRA TIB LT SZ3 ALL POLY ULTRA TIB RT SZ3 ALL POLY ULTRA TIB RT SZ5

Reason

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Identifiers

code_infoItem Number 621800809 621801809 621801819 621801913 621801916 621810816 621810909 621820809 621820811 621820911 621830811 621830909 621850916

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1262-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.