FDA Device recall Z-1259-2018

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System…

On 2018-04-04 the FDA classified a Class II recall of Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System… by Encore Medical, citing it was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-50….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1259-2018
ClassificationClass II
Statuscompleted
Initiated2018-01-29
Published2018-04-04
Last updated2026-09-13 11:54 UTC
CompanyEncore Medical
DistributionUS

Product

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

Reason

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

Identifiers

code_infoAll Serial Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: completed.