FDA Device recall Z-1258-2022

t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology

On 2022-06-22 the FDA classified a Class II recall of t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology by Tandem Diabetes Care, citing insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to le….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1258-2022
ClassificationClass II
Statusongoing
Initiated2020-09-10
Published2022-06-22
Last updated2026-09-13 11:56 UTC
CompanyTandem Diabetes Care
Distributionn/a

Product

t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology

Reason

Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stop, which could lead to hyperglycemia.

Identifiers

code_infoDexcom G5 - All Serial Numbers; Basal-IQ Technology - Software Versions 6.3.01 and Earlier