FDA Device recall Z-1258-2018
Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460 Product Usage: The AltiVate Anatomic Shoulder System…
- FDA Device
- Class II
- completed
- Published 2018-04-04
On 2018-04-04 the FDA classified a Class II recall of Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460 Product Usage: The AltiVate Anatomic Shoulder System… by Encore Medical, citing it was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-50….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1258-2018 |
| Classification | Class II |
| Status | completed |
| Initiated | 2018-01-29 |
| Published | 2018-04-04 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Encore Medical |
| Distribution | US |
Product
Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
Reason
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Identifiers
| code_info | All Serial Numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: completed.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1258-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.