FDA Device recall Z-1256-2022
Reliance Synergy Washer/Disinfector
- FDA Device
- Class II
- terminated
- Published 2022-06-15
On 2022-06-15 the FDA classified a Class II recall of Reliance Synergy Washer/Disinfector by Steris, citing the firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could ove….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1256-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-04-22 |
| Published | 2022-06-15 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Steris |
| Distribution | US |
Product
Reliance Synergy Washer/Disinfector
Reason
The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Identifiers
| code_info | UDI-DI: 00724995152161 Serial Numbers: 3633702001 3633702002 3633802007 3633702004 3633902002 3602303011 3602…UDI-DI: 00724995152161 Serial Numbers: 3633702001 3633702002 3633802007 3633702004 3633902002 3602303011 3602403002 3604403007 3605203010 3603003006 3603403007 3604403006 3633802008 3606503002 3606403002 3606403003 3606503001 3605103003 3607303014 3605203011 3607303013 3633302004 3610703003 3610803004 3609203001 3609403001 3609003005 3633702005 3610803002 3612103005 3615703001 3615703002 3612203006 3612703013 3615303012 3602303013 3615003008 3604503002 3614003006 3618803002 3618803003 3614603009 3618203003 3620903009 3608503002 3614803010 3614803012 3614803011 3622403010 3622303020 3609003004 3609203002 3623703010 3621803008 3614603010 3614903008 3623703011 3623703012 3624503005 3620903011 3624503006 3628103009 3629303002 3629303005 3630203001 3630303002 3631703001 3628103008 3628103010 3621603009 3622603008 3622003002 3621703005 3626603004 3626903003 3626903004 3623703009 3623403010 3634903010 362890301 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1256-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.