RecallRadar

FDA Device recall Z-1253-2023

In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116

On 2023-03-22 the FDA classified a Class II recall of In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116 by Biofire Diagnostics, citing due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature e….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1253-2023
ClassificationClass II
Statusongoing
Initiated2023-02-01
Published2023-03-22
Last updated2026-09-13 11:56 UTC
CompanyBiofire Diagnostics
DistributionUS

Product

In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116

Reason

Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.

Identifiers

code_infoPart Number: RFIT-ASY-0116 UDI: 00815381020109 Pouch Lot Number: 1674322

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1253-2023%22