RecallRadar

FDA Device recall Z-1251-2021

McKesson PREVENT M NEEDLE (Magellan Technology) 25G X 5/8". A single lumen needle intended to inject fluids into, or withdraw fluids from,…

On 2021-03-24 the FDA classified a Class II recall of McKesson PREVENT M NEEDLE (Magellan Technology) 25G X 5/8". A single lumen needle intended to inject fluids into, or withdraw fluids from,… by Cardinal Health 200, citing safety shield not locking into the cannula. If the safety shield activation fails to lock a contaminated needle, a needlestick may occur and there is the potential for exposure to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1251-2021
ClassificationClass II
Statusterminated
Initiated2021-02-09
Published2021-03-24
Last updated2026-09-13 11:56 UTC
CompanyCardinal Health 200
DistributionUS

Product

McKesson PREVENT M NEEDLE (Magellan Technology) 25G X 5/8". A single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. Product Code: 26558

Reason

Safety shield not locking into the cannula. If the safety shield activation fails to lock a contaminated needle, a needlestick may occur and there is the potential for exposure to bloodborne pathogens

Identifiers

code_infoLot Code: 028603