FDA Device recall Z-1250-2014

MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX…

On 2014-03-26 the FDA classified a Class II recall of MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX… by Zimmer, citing A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was adva….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1250-2014
ClassificationClass II
Statusterminated
Initiated2014-02-25
Published2014-03-26
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX TIB INS 6MM

Reason

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Identifiers

code_infoItem Number 502001016 502004006 502004009 502004016 502010006