FDA Device recall Z-1249-2023
Plato 17 Microcatheter: Straight Tip, PL17-160-000
- FDA Device
- Class II
- ongoing
- Published 2023-03-22
On 2023-03-22 the FDA classified a Class II recall of Plato 17 Microcatheter: Straight Tip, PL17-160-000 by Scientia Vascular, citing microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1249-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-02-01 |
| Published | 2023-03-22 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Scientia Vascular |
| Distribution | US |
Product
Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090
Reason
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
Identifiers
| code_info | All lots, REF/UDI-DI/Lot(Expiration): PL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2…All lots, REF/UDI-DI/Lot(Expiration): PL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2022), 900177(9/30/2022), 900180(9/30/2022), 900186(10/31/2022), 900187(10/31/2022), 900188(10/31/2022), 900193(10/31/2022), 900196(12/31/2022), 900199(12/31/2022), 900202(12/31/2022), 900221(2/28/2023), 900222(2/28/2023), 900225(2/28/2023), 900226(3/31/2023), 20441(11/30/2023), 20476(11/30/2023), 20650(12/31/2023); PL17-160-045/00818075010462/900167(8/31/2022), 900168(8/31/2022), 900169(8/31/2022), 900174(8/31/2022), 900175(8/31/2022), 900179(9/30/2022), 900181(9/30/2022), 900183(10/31/2022), 900185(10/31/2022), 900205(1/31/2023), 900206(1/31/2023), 900210(1/31/2023), 900211(1/31/2023), 900215(1/31/2023), 900220(1/31/2023), 900219(1/31/2023), 900217(2/28/2023), 900223(2/28/2023), 900224(2/28/2023); PL17-160-090/00818075010479/900170(8/31/2022), 900171(8/31/2022), 900178(9/30/2022), 900182(9/30/2022), 900184(10/31/2022), 900192(11/30/2022), 900195(11/30/2022), 900209 |
|---|