FDA Device recall Z-1249-2014
Natural-Knee¿ N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1
- FDA Device
- Class II
- terminated
- Published 2014-03-26
On 2014-03-26 the FDA classified a Class II recall of Natural-Knee¿ N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1 by Zimmer, citing A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was adva….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1249-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-02-25 |
| Published | 2014-03-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
Natural-Knee¿ N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1
Reason
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Identifiers
| code_info | Item Number 00591101128 00591101137 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1249-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.