FDA Device recall Z-1248-2022
Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and dryin…
- FDA Device
- Class II
- ongoing
- Published 2022-06-15
On 2022-06-15 the FDA classified a Class II recall of Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and dryin… by Getinge Usa Sales, citing verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1248-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-04-13 |
| Published | 2022-06-15 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Getinge Usa Sales |
| Distribution | US |
Product
Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, etc. Model: CM320 WU
Reason
Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection
Identifiers
| code_info | UDI-DI: 07340153700185 Serial Numbers: W50034570 WAA070170 WAA070618 WAA081312 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1248-2022%22