FDA Device recall Z-1241-2025
Optima MR360, NMRI system
- FDA Device
- Class II
- ongoing
- Published 2025-03-05
On 2025-03-05 the FDA classified a Class II recall of Optima MR360, NMRI system by Ge Healthcare China, citing for certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1241-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-01-31 |
| Published | 2025-03-05 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Ge Healthcare China |
| Distribution | US |
Product
Optima MR360, NMRI system
Reason
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Identifiers
| code_info | A. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI System IDs 082427040134 082427040185 082427060043 08242707…A. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI System IDs 082427040134 082427040185 082427060043 082427070088 082427070093 082427070099 082427100174 082427100178 082427100197 082427120140 082427140095 082427220046 082427270041 082427310088 082427310092 083027288790310 0856270061 AE1373MR01 DZ1339MR03 FI1432MR01 H4678MR04 MRR9282 MRR9877 PL2059MR01 PL2433MR01 RU3338MR03 SA1056MR03 SA2191MR04 SA2253MR01 SA2352MR01 TN1008MR01 TN1241MR01 34352MRS01 (Update 12/29/25 - Additional System IDs 83027809699615, LV4026MR01) (Update 1/28/26 - Additional System ID 82427190104) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.