FDA Device recall Z-1240-2016

Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic u…

On 2016-03-30 the FDA classified a Class II recall of Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic u… by Siemens Healthcare Diagnostics, citing the drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1240-2016
ClassificationClass II
Statusterminated
Initiated2016-01-13
Published2016-03-30
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur¿, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.

Reason

The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

Identifiers

code_infoAll lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.