FDA Device recall Z-1235-2023

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2

On 2023-03-15 the FDA classified a Class II recall of Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2 by Sensus Healthcare, citing when the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1235-2023
ClassificationClass II
Statusterminated
Initiated2023-02-10
Published2023-03-15
Last updated2026-09-13 11:56 UTC
CompanySensus Healthcare
DistributionFL, IA, IL, IN, NC, NM, NY

Product

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

Reason

When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.

Identifiers

code_info
Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 16…Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272