RecallRadar

FDA Device recall Z-1234-2026

EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H

On 2026-02-11 the FDA classified a Class II recall of EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H by Encore Medical, citing 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1234-2026
ClassificationClass II
Statusongoing
Initiated2026-01-05
Published2026-02-11
Last updated2026-09-13 11:58 UTC
CompanyEncore Medical
DistributionAL, CA, IA, MA, MI, MO, MS, NJ, NY, OK, PA, PR, RI, TX

Product

EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C

Reason

942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

Identifiers

code_info
REF/UDI-DI/Lot(Expiration): 942-01-36H/00190446309789/624Z1032(03-15-2026), 942-01-36J/00190446309802/626Z1021(03-15-2…REF/UDI-DI/Lot(Expiration): 942-01-36H/00190446309789/624Z1032(03-15-2026), 942-01-36J/00190446309802/626Z1021(03-15-2026), 942-01-40G/00190446309826/628Z1034(03-15-2026), 942-01-32C/00190446309697/615Z1026(03-15-2026)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1234-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.