RecallRadar

FDA Device recall Z-1232-2017

Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment

On 2017-03-01 the FDA classified a Class II recall of Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1232-2017
ClassificationClass II
Statusterminated
Initiated2015-01-05
Published2017-03-01
Last updated2026-09-13 11:54 UTC
CompanyTerumo Cardiovascular Systems
Distributionn/a

Product

Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment

Reason

Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.

Identifiers

code_info
Catalog Number: 401-231, Product Description: Titan" Stabilizer Attachment , Lot Number: 052R Catalog Number 401-2…Catalog Number: 401-231, Product Description: Titan" Stabilizer Attachment , Lot Number: 052R Catalog Number 401-231u , Product Description TitanTM 360 Stabilizer Attachment , Lot Number: 020R, 024R, 026R, 029R, 030R,034R,035R,037R, 040 , 048R

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1232-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.