RecallRadar

FDA Device recall Z-1231-2025

BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the conti…

On 2025-03-05 the FDA classified a Class II recall of BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the conti… by Carefusion 303, citing due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could resul….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1231-2025
ClassificationClass II
Statusongoing
Initiated2025-01-23
Published2025-03-05
Last updated2026-09-13 11:57 UTC
CompanyCarefusion 303
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY

Product

BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration

Reason

Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.

Identifiers

code_infoCatalog Number: 8100 UDI-DI codes: 10885403810046 10885403517723 Software Version: 12.1.2 Serial/Lot Numbers: Pending