RecallRadar

FDA Device recall Z-1229-2023

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-…

On 2023-03-15 the FDA classified a Class II recall of Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-… by Datascope, citing there have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1229-2023
ClassificationClass II
Statusongoing
Initiated2023-02-07
Published2023-03-15
Last updated2026-09-13 11:56 UTC
CompanyDatascope
DistributionUS

Product

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, and 0998-00-0800-55

Reason

There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.

Identifiers

code_info
All Serial Numbers. Model Number: 0998-00-0800-31, UDI-DI:10607567109053; Model Number: 0998-00-0800-32, UDI-DI: 1060…All Serial Numbers. Model Number: 0998-00-0800-31, UDI-DI:10607567109053; Model Number: 0998-00-0800-32, UDI-DI: 10607567111117; Model Number: 0998-00-0800-33, UDI-DI: 10607567109008; Model Number: 0998-00-0800-34, UDI-DI: 10607567111940; Model Number: 0998-00-0800-35, UDI-DI: 10607567109107; Model Number: 0998-00-0800-45, UDI-DI: 10607567108421; Model Number: 0998-00-0800-52, UDI-DI: 10607567108438; Model Number: 0998-00-0800-53, UDI-DI: 10607567108391; Model Number: 0998-00-0800-55, UDI-DI: 10607567108414; Model Number: 0998-00-0800-65, UDI-DI: 10607567113432

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1229-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.