RecallRadar

FDA Device recall Z-1224-2013

Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish

On 2013-05-15 the FDA classified a Class II recall of Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish by Stryker Endoscopy, citing the pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1224-2013
ClassificationClass II
Statusterminated
Initiated2013-03-27
Published2013-05-15
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Reason

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Identifiers

code_infoManufacturer Part Number: 250-070-411

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1224-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.